Tecfluid MX4 For FLOMID Series Manuel d'utilisateur Page 1

Naviguer en ligne ou télécharger Manuel d'utilisateur pour Capteurs Tecfluid MX4 For FLOMID Series. Tecfluid MX4 For FLOMID Series User Manual Manuel d'utilisatio

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer

Résumé du contenu

Page 1 - Instructions manual

Sensor FLOMID-FX Converter MX4 Series Flomid Instructions manual R-MI-FIMX4 Rev.: 1 English version

Page 2

10 4.5 Aeration If there is a point where the difference in level is higher than 5 m an air inlet valve should be installed after the sensor to avoid

Page 3

11 4.8 Magnetic fields Strong magnetic fields close to the sensor should be avoided. 4.9 Temperature In open air installations it is recommended to

Page 4

12 5 MOUNTING 5.1 Parallelism The maximum parallelism error must be less than 0.5 mm (Lmax—Lmin ≤ 0.5 mm). 5.2 Gasket position For the wafer mounted

Page 5

13 b) In the case of metallic pipes with internal lining or plastic pipes, connect the earth cables to the earthing disks, supplied on request. 5.3

Page 6 - V = B·v

14 Earthing disks are supplied in two versions: METALLIC, disk in stainless steel EN 1.4404 (SS 316L), for liquids compatible with this material. P

Page 7 - 3 HANDLING

15 MX4 CONVERTER 1 INTRODUCTION The MX4 converter unit can be used with the different FLOMID and FLOMAT series of electromagnetic flow sensors. The e

Page 8

16 2.1.2 Remote converter One of the ends of the cable has a header and two wires, to connect them to the sensor. The connection is as explained in p

Page 9

17 2.2.1 Power supply wiring Before starting the installation of the equipment, check that the supply voltage available is the same as marked on the

Page 10

18 Terminal Description Relay 6 Normally Open Relay 1 5 Common Relay 1 4 Normally Open Relay 2 3 Common Relay 2 The relay ou

Page 11

19 2 Terminal 1 Contact 2 Contact These terminals correspond to an input that resets the value of the totalizer. A potential free normally open

Page 12 - 5 MOUNTING

2 Thank you for choosing the flowmeter series FLOMID from Tecfluid S.A. This instruction manual allows the installation and operation of the electro

Page 13

20 2.2.5 Pulse output wiring Terminal 14 Collector 15 Emitter The pulse output is opto-isolated. The terminals are the collector and emitter of

Page 14 - 6 MAINTENANCE

21 Example. Connection with the load at the collector Example. Connection with the load at the emitter 3 REMOTE SENSOR When an installation requires

Page 15 - MX4 CONVERTER

22 REMOTE SENSOR 3

Page 16

23 3.2 Cable installation The following points must be taken into account:  The cable should be installed in a conduit or should be securely fixe

Page 17

24 Terminal 16 Top coil (red cable) 17 Bottom coil (black cable) 18 Rear electrode (white cable) 19 Ground (shield) 20 Front electrode

Page 18

25 3.5 Cable specifications Model: CERVITRONIC PAR-POS Code 04754502 Construction Conductor: Annealed electrolytic copper, tinned As Norm:

Page 19

26 The following figure shows the functionality of the converter keys. (Down / Left) To switch between flow rate, totalizer and fluid velocity screen

Page 20 - Passive output

27 To change the access password of the programming menu, select "Programming" in the main menu and then "Password". 5 MAIN MENU

Page 21 - 3 REMOTE SENSOR

28 5 MAIN MENU When the "Password" option is selected, a screen that indicates the password status for this menu appears. Selecting “enabl

Page 22 - REMOTE SENSOR

29 Once the password is entered, and after pressing the key (Enter), the first screen allows to choose between the different options of the installat

Page 23 - Connector cover

3 TABLE OF CONTENTS FLOMID-FX SENSOR 1 WORKING PRINCIPLE ... 6 2 RECEPTI

Page 24

30 6.5 Diagnosis When the flowmeter is not working properly, it can be useful to test some parts of the instrument in order to locate the problem.

Page 25

31 6.5.1 Parameters This screen allows the diagnosis of the coil current, the differential voltage on the sensor electrodes, and the conductivity of

Page 26 - 4 CONVERTER INTERFACE

32 Press again the Escape key until the working screen. 6.5.3 Mains frequency If the power supply of the converter is AC, the following screen will a

Page 27 - 5 MAIN MENU

33 7 CONVERTER PROGRAMMING By programming the converter the visualization and the outputs of the instrument can be configured. Turn on the converter

Page 28

34 7.2 Units In this screen the units for the liquid velocity, the flow rate and the totalizer can be chosen independently. 7.3 Flow rate decimals I

Page 29

35 7.4 Flow rate This screen displays a submenu to modify different aspects related to flow reading, such as empty pipe detection, cut off, reversal

Page 30

36 7.4.3 Reversal flow rate The MX4 is a bi-directional converter. The flow rate changes its sign automatically when the flow is reversed. If after

Page 31

37 7.4.5 Offset In order to obtain a perfect linearization of the instrument, it is recommended to make an adjustment of the zero offset each time an

Page 32

38 Example If the activation point is programmed to 100 m3/h and the deactivation point is programmed to 90 m3/h, when the flow rate is zero the outp

Page 33 - 7 CONVERTER PROGRAMMING

39 If “Unidirectional” is chosen, the output will give pulses proportional to flow rate when the flow rate is positive, and will not give them when t

Page 34

4 3 REMOTE SENSOR ... 21 3.1 Preparing the cable ...

Page 35 - 7.4.2 Cut off

40 7.5.3.1 4-20 mA output configuration It allows to choose between unidirectional and bidirectional. Programming a flow rate value for 20 mA higher

Page 36

41 In this screen you can choose the flow direction for which you want to configure the action that the totalizer will do. Positive flow is defined a

Page 37

42 10 WORKING SCREEN When exiting the menu, the display shows the default screen. To scroll between the three operating screens, press the key (Down

Page 38

43 12 MAINTENANCE No special maintenance is required. For cleaning, a humid cloth can be used, and if necessary with a little neutral soap. Solvents

Page 39

44 13 ASSOCIATED SOFTWARE WINSMETER MX4 Most of the steps in the preceding paragraphs can be done by means of the device associated software Winsmet

Page 40

45 13.2 Port connection In the "Port" section, choose the appropriate port for the converter. This will appear with the name of the port fo

Page 41

46 13.3 Access to installation and programming In order to change the data contained in the "Install" tab, you must enter a password. The

Page 42 - 8 SERIAL NUMBER

47 In the installation window the parameters that adapts the converter to a sensor and to a determinated installation can be configured. To transfer

Page 43 - 11 STARTING CURRENT

48 Likewise, to enter into the programming window, just click the corresponding tab. Changing the parameters of this screen, (see previous page) you

Page 44

49 ASSOCIATED SOFTWARE WINSMETER MX4 13.5 Firmware updates New firmware updates can be published in the website. These updates contain improvements o

Page 45

5 7.5.2 Pulse output ... 38 7.5.2.1 Pulse output configuration ...

Page 46

50 14 TECHNICAL CHARACTERISTICS Accuracy reading value for v > 0,4 m/s reading value for v < 0,4 m/s Repeatabili

Page 47

51 14 Flow rate indication Nº of digits: 5 (up to 2 decimals configurable)** Digit size: 11 mm ** When the available digits are full and the i

Page 48

52 15 DIMENSIONS 15 DIMENSIONS FLOMID-0FX (EN 1092-1 wafer mounted) FLOMID-0FX (ANSI B16.5 wafer mounted) DN g L F A Ax h Weight (kg)

Page 49

53 15 DIMENSIONS

Page 50

54 FLOMID sensor flanged mounted, common dimensions DN ANSI L1 L2 A Ax h Weight (kg) 10 ⅜" 85 150 290 304 202 4.5 1

Page 51

55 DIMENSIONS 15

Page 52 - (EN 1092-1 wafer mounted)

56 DN L1 L2 D A Ax h F 10 104 120 60 260 274 172 48 15 104 120 65 260 274 172 48 20 104 120 70 262 276 174 5

Page 53 - DIMENSIONS

57 FLOMID-1FX (DIN 11851) DN di R D L1 L2 L3 H 10 10 Rd 28 x 1/8" 60 104 120 200 48 15 16 Rd 34 x 1/8"

Page 54 - FLOMID-4FX

58 DIMENSIONS FLOMID-5IFX (CLAMP ISO 2852) DN OD* di C D L1 L2 L3 H 10 12.0 10.0 34.0 60 104 120 200 48 10 12.7 10.7

Page 55

59 DIMENSIONS FLOMID-5TFX (TRI-CLAMP®) DN Coup. OD* di C D L1 L2 L3 H 10 ½" 12.7 9.4 25.0 60 104 120 200 48 15 ¾&quo

Page 56

6 1 2 1 WORKING PRINCIPLE The FLOMID electromagnetic flowmeters are based on Faraday’s induction law. When an electrically conductive liquid flows t

Page 57 - FLOMID-3FX (SMS 1145)

60 DIMENSIONS FLOMID-7FX (ISO 2853) DN di R C D L1 L2 L3 H 10 10.0 Tr 22.89 x 1/8" 15.0 60 104 120 200 48

Page 58 - FLOMID-5IFX (CLAMP ISO 2852)

61 DIMENSIONS FLOMID-9WDFX (DIN 11850 weld-on connections) DN de* di D L1 L2 L3 H 10 13 10 60 104 120 180 38 15 19 16 65

Page 59 - FLOMID-5TFX (TRI-CLAMP®)

62 16 TROUBLESHOOTING 16 Problem Probable cause Solution The screen shows empty pipe Pipe is empty Make sure that the pipe is completely full, for

Page 61

64 WARRANTY Tecfluid S.A. guarantees all the products for a period of 24 months from their sale, against all faulty materials, manufacturing or perfo

Page 62 - 16 TROUBLESHOOTING

7 4 INSTALLATION This should be made in a point that guarantees that the pipe is always completely full. Avoid high points of the pipes where air poc

Page 63 - ANNEX A

8 4.2 Straight pipe sections They are necessary before and after the sensor. The minimum distances are the following: Upstream 5 DN Downstream

Page 64

9 4.2.1 Mixtures If liquids of different conductivities are mixed it is necessary to install the sensor a minimum of 30 DN from the point of mixture

Commentaires sur ces manuels

Rosman Mohamed 12 Jul 2024 | 01:50:59

thank you