Tecfluid CU Series Manuel d'utilisateur Page 22

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 48
  • Table des matières
  • DEPANNAGE
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 21
22
5.11 Installation and distance between the transducers
This screen shows the theoretical distance (Dt) at which the transducers should be placed
to work in the best conditions. The practical distance (Dp) is explained in the paragraph
5.14.
It is recommended to fix initially only one of the transducers since in general there are
small differences between the theoretical and the optimal distance. The fine adjustment of
the position of the second transducer is done by using the option "Signal Strength" of the
menu (see paragraph 5.13 page 24).
In case the installation is made by means of the alignment guide, it is recommended to
place the transducer TR1 with the gel at the end of the guide so that its inner side
matches the zero point of the ruler.
Transducer
Fastening “U”
Inner side
To place the transducers it is essential to apply a layer of ultrasound gel at its base. This
gel is supplied with the equipment.
5.12 Fixing the transducers
The transducers should be placed so that the liquid flows from TR1 to TR2, in order that
when the liquid flows in the usual direction, the displayed flow rate has a positive sign.
5
TRANSDUCERS INSTALLATION
V mode
Z mode
Vue de la page 21
1 2 ... 17 18 19 20 21 22 23 24 25 26 27 ... 47 48

Commentaires sur ces manuels

Pas de commentaire