Tecfluid PT Series Manuel d'utilisateur

Naviguer en ligne ou télécharger Manuel d'utilisateur pour Instruments de mesure Tecfluid PT Series. Tecfluid PT Series User Manual Manuel d'utilisatio

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 4
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 0
Conforms with the Pressure
Equipment Directive 97/23/CE.
This equipment is considered as being a
pressure accessory and NOT a safety
accessory as defined in the 97/23/CE
directive, Article 1, paragraph 2.1.3.
PT / PS / PTM / PSM
Variable area
flowmeters
Series
Instructions Manual
0830
!
C-MI-11 Rev.: 1
English Version
instrumentation
for fluids
Technical Data
Accuracy: Based on VDE / VDI 3513
- PT/PS ± 4% of full scale value
- PTM/PSM ± 6% of full scale value
Material:
- PT/PTM Trogamid T
- PS/PSM: Polysulfphone
Mounting: Vertical (Rising flow direction)
Pipe fittings:
- PT-11/PS-31 BSP female thread
- PT-12/PS-32 DIN flanges
- PTM/PSM-01 BSP female thread
- PTM/PSM-02 DIN flanges
Max. Working pressure:
from 8 to15 bar at 20 ºC depending on the tube
Max.Fluid temperature: 0 ... +60ºC (Trogamid T)
0 ... +90ºC (Polysulphone)
Vue de la page 0
1 2 3 4

Résumé du contenu

Page 1 - Instructions Manual

• Conforms with the Pressure Equipment Directive 97/23/CE. This equipment is considered as being a pressure accessory and NOT a safety accessory

Page 2 - Reading line

Working principle The flowmeter consists of a float inside a conical tube. The rising flow pushes the float to an equilibrium point. The area obtained

Page 3

3 CLEANING AND MAINTENANCE For routine cleaning, normally it is sufficient to pass a solution of water and detergent through the flowmeter. If the flo

Page 4

The float must be cleaned with a soft brush, NEVER WITH METALLIC UTENSILS. To assemble the unit follow the disassembly process in reverse. Insert the

Modèles reliés PS Series

Commentaires sur ces manuels